VL-2000 Relative Humidity and Temperature Data Logger, humidity data logger, humidity data loggers, validatable humidity data loggers with temperature sensors, Temperature and Humidity data loggers

Request pricing & availability

2000 Humidity Recorder Datasheet

Related Information

Veriteq Validation System Overview

Validation System Components

vLog Software

Common questions

Validation protocol

Sample Documents

Calibration Services

Industry Solutions

Validation Data Logger Rentals

Articles & White Papers

Veriteq's Solution for 21 CFR Part 11 Compliance

"Catching the Drift: What the specifications of your Humidity Measurement Device might be missing"
- published in Cal Lab Magazine

Solutions

Clinical Trial Refrigerator & Freezer validation

Pharmaceutical chamber validation

Stability Studies

Pharmaceutical warehouse validation

Veriteq loggers used to help with large Biotech company relocation



Validatable Relative Humidity & Temperature Data Loggers - Specifications


Veriteq's VL 2000 series loggers are fully validatable and designed for regulated industries/pharmaceutical applications — with NIST-traceable, ISO/IEC 17025 calibration, cGMP/FDA compliant measurement and redundant recording of temperature and humidity.

  • Portable self-contained relative humidity & temperature data logger for critical validation applications

  • Equipped with internal relative humidity & temperature sensors, memory and ten-year battery

  • Optional external channel for voltage or current inputs allow recording of differential pressure, level, particles, conductivity or C02

  • Optional Boolean channels for door switches or alarm contacts

  • Tamper-proof operation produces secure documentation for validation records

  • Password-protected setup, download and calibration functions ensure integrity of collected data

  • Records relative humidity (RH) and temperature without wires, power cords or connections

  • Traceable instrument-grade RH and temperature measurement accuracy

  • Adjustable logging intervals with multi-year data recording capacity

  • Premium quality precision sensors offer exceptional in-calibration data logger performance

Technical Specifications

Internal Relative Humidity Sensor

Type of sensor Integrated high-accuracy capacitive polymer RH sensor
Initial Accuracy

±1%RH over 10%RH to 90%RH at +20ºC to +30ºC

±2%RH over 10%RH to 90%RH at -20ºC to +70ºC

Operating range -35°C to +85°C, 0% RH to 100% RH (non-condensing)
Resolution 0.05% RH
1-year Accuracy

(*denotes Accuracy One Year after original deployment)

*±2% RH over 10% RH to 90% RH at +20°C to +30°C
 
±3% RH over 10% RH to 90% RH at -20°C to +70°C
Calibration Interval Annual (recommended)

Internal Temperature Sensor

Initial Accuracy ±0.10ºC over +20ºC to +30ºC

±0.15ºC over -25ºC to +70ºC
Operating range -35 to +85°C (-31 to 185°F)
Resolution 0.02°C at +25°C
1-year Accuracy ±0.15°C over 20°C to +30°C
±0.25°C over -25°C to +70°C

General Humidity Logger Specifications

Size 3.4 x 2.3 x 1" (85x59x26mm);
76g (2.7 oz)
Operating Range -35 to 85°C (-31 to 185°F)
and 0% RH to 100% RH
Interfaces RS-232 serial, USB, Ethernet, WiFi, PoE network interface available.
Mounting Magnetic strips; 3M Dual Lock™ fasteners
Software Compatible with the following software packages: Veriteq Spectrum, Veriteq viewLinc (monitoring & alarming applications), Spectrum Excel Add-in, Spectrum API (for custom applications) and OPC Server to add loggers onto compatible monitoring systems.
Clock Accuracy: ± 1 min./month at -25° C to 70°C.
Electromagnetic Compatibility

FCC Part 15 and CE
EN55022:2006
EN61000-4-2:2001
EN 61000-4-3:2006

Power Source Internal lithium battery with life of 10 years at 1 min. sampling rate or longer. See note on sample rates and battery life.

Memory

Memory Type Non-volatile EEROM
Data Sample Capacity 122,197 12-bit samples
Memory Modes User-selectable wrap (FIFO) or stop when memory is full. User-selectable start and stop times.
Memory Protection Data retention > 20 years backup without power.
Sampling Rates User-selectable (in intervals of 10 seconds) from once every 10 seconds to once a day. Note: 10-year battery life is specified at 1 minute sample rate. Shorter sample intervals will affect battery life.

Channel Configuration & Recording Span

Model Number
Channel Types
CH1
CH2
CH3
CH4
2000-3CR
Temperature
Relative Humidity
Current 4 to 20 mA
2000-35R
Temperature
Relative Humidity
Voltage 0 to 5 VDC
2000-3AR
Temperature
Relative Humidity
Voltage 0  to 10 VDC
2000-4BR
Temperature
Relative Humidity
Boolean
Boolean
Sample Interval
10 seconds
14.1 days
7.1 days
4.7 days
3.5 days
1 Minute
2.8 months
1.4 months
23.8 days
21.2 days
5 Minutes
1.2 years
7.1 months
4.7 months
3.5 months
15 Minutes
3.5 years
1.7 months
1.2 years
10.6 months
1 hour
13.9 years
7.0 years
4.6 years
3.5 years

Current Loop & Voltage Inputs

Input Type

Current Loop

Analog Voltage

Available Ranges
0 to 22mA
0 to 5VDC, 0 to 10VDC
Resolution
5.5μA
0.025%F.S.
Accuracy
±0.15% at +25°C (+77°F)
±0.15% at +25°C (+77°F)
Input Impedances
>1MOhms
Isolation
One common for logger
One common for logger
Overload Protection
40 mA max. (reverse-polarity protected)
±24 VDC max. (reverse-polarity protected)

 

*Temperature channel must be enabled when the RH channel is enabled. **Termination resistance plus approximately 0.4 volt drop through a protection diode.